3 stdadd dos - standard addition – Metrohm tiamo 2.3 (process analysis) User Manual

Page 1091

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.3 (for Process Analysis)

■■■■■■■■

1075

automatic
If a temperature sensor is connected then the temperature will be
measured continuously. Otherwise, the temperature entered manually
under Temperature on the Measuring parameters tab will be
used.
off
The temperature will not be measured. The temperature manually
entered under Temperature on the Measuring parameters tab will
be used.

Stirrer

Stirrer

Selection of the stirrer.

Selection

1 | 2 | 3 | 4 | off

Default value

1

off
Means that no stirrer will be used.

Stirring rate

Setting the stirring rate. The algebraic sign of the stirring rate changes the
direction in which the stirring is done.

Range

-15 ... 15

Default value

8

Switch off automatically

on | off (Default value: on)

If this option is enabled, the stirrer will be switched off automatically
when the command has finished.

5.6.4.18.3

STDADD dos - Standard addition

Tab: Method

STDADD dos Properties... Standard addition

Command name

Name of the command.

Input

25 characters

The parameters for the standard addition can be defined on this tab

Number of additions

Number of times the standard solution is added.

Range

1 ... 19

Default value

3

Advertising