5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

530

■■■■■■■■

tiamo 2.3 (for Process Analysis)

(Measured value, Time or Volume) for the fixed endpoint from the
measuring point list.

Moves the selected line up (changes the sequence).

Moves the selected line down (changes the sequence).

[New]

Open the dialog window Fixed endpoint evaluation # to enter a new
fixed endpoint (see Chapter 5.6.3.2.4.11, page 537).

[Properties]

Open the dialog window Fixed endpoint evaluation # to edit the fixed
endpoint selected in the table (see Chapter 5.6.3.2.4.11, page 537).

[Delete]

Delete the selected line.

Minimum evaluation

Minimum evaluation

on | off (Default value: off)

If this option is enabled, then the associated volume, time and tempera-
ture for the minimum measured value will be interpolated from the meas-
uring point list.

Threshold value

The evaluation of the minimum begins as soon as the slope of the curve
exceeds the set threshold value.

Range

1.0 ... 2000.0 mV/mL

Default value

25.0 mV/mL

Maximum evaluation

Maximum evaluation

on | off (Default value: off)

If this option is enabled, then the associated volume, time and tempera-
ture for the maximum measured value will be interpolated from the meas-
uring point list.

Threshold value

The evaluation of the maximum begins as soon as the slope of the curve
exceeds the set threshold value.