Metrohm tiamo 2.3 (process analysis) User Manual

Page 1079

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.3 (for Process Analysis)

■■■■■■■■

1063

Measuring interval

Time interval for entering a measured value in the measuring point list.

Range

0.25 ... 999999.00 s (Increment: 0.25)

Default value

2.00 s

Stop measured value

Stop when the entered measured value has been reached since the start
of the measurement.

Range

0 ... 4000 mAU

Selection

off

Default value

off

off
off
means that no stop will take place.

Temperature

Temperature

Measuring temperature which can be entered manually. If a temperature
sensor is connected and the Temperature measurement is set on the
General/Hardware tab under Sensor to automatic or continuous,
then the temperature will be measured continuously.

Range

20.0 ... 60 °C

Default value

25.0 °C

Measurement without drift control

Duration of measurement

Duration of the adsorption measurement.

Range

0 ... 999999 s

Default value

120 s

Measuring interval

Time interval for entering a measuring point in the measuring point list.

Range

0.25 ... 999999.00 s (Increment: 0.25)

Default value

2.00 s

Stop measured value

Stop when the entered measured value has been reached since the start
of the measurement.

Range

0 ... 4000 mAU

Selection

off

Default value

off