5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

904

■■■■■■■■

tiamo 2.3 (for Process Analysis)

Wait for [Continue]
Reagent dosing in the running STAT command will be interrupted and
a message will be displayed. As soon as the monitored measured value
is again within the limits (including hysteresis) reagent dosing can be
resumed by pressing the [Continue] button in this message box.
Wait for limit ok
Reagent dosing in the current STAT command will be interrupted. As
soon as the monitored measured value is again within the limits
(including hysteresis), reagent dosing will be resumed automatically.
none
No action will be taken if limits are exceeded.

Monitoring Dosing rate

Monitoring Dosing rate

on | off (Default value: off)

If this check box is activated, then the averaged dosing rate will be moni-
tored and any exceeded limits will be entered in the measuring point list.
The monitoring of the dosing rate is not started until 10 s after the start of
the titration.

Lower limit

Lower limit of the dosing rate. If the dosing rate falls below this limit, the
event Dosing rate lower limit exceeded is triggered.

Titrando, 855
Range

0.00 ... 166.00 mL/min

Default value

0.00 mL/min

Titrino
Range

0.00 ... 150.00 mL/min

Default value

0.00 mL/min

Lower hysteresis

Lower hysteresis of the dosing rate. If the dosing rate exceeds the lower
limit by this hysteresis value, then the event Dosing rate lower limit ok
is triggered.

Titrando, 855
Range

0.00 ... 166.00 mL/min

Default value

0.20 mL/min

Upper limit

Upper limit of the dosing rate. If the dosing rate exceeds this limit, the
event Dosing rate upper limit exceeded is triggered.