5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

716

■■■■■■■■

tiamo 2.3 (for Process Analysis)

Titration parameters
Parameters for the run of the titration.

Stop conditions
Definitions of conditions which cause the stop of the titration.

Conditioning
Enable/Disable the conditioning. Definition of the start drift, drift cor-
rection and stop conditions.

Additional evaluations
Definition of additional methods for the evaluation of the titration
curves.

Additional measured values
Definition of additional measured values of other measuring com-
mands which are saved as additional columns in the measuring point
list.

Command variables
The following command variables are generated in the method run by the
command SET U and can be used in formulas under the designation
'Command name.Variable designation':

Designation

Description

.BLV

Blank value of the sensor used for the command (only
for ISE sensors)

.BSY

Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY
; invalid (variable not available) = Command
has never been started

.CBY

Command status; 1 = Conditioning activated, 0 =
Conditioning deactivated

.COK

Command status; 1 = Condition requirement fulfil-
led
, 0 = Condition requirement not fulfilled

.CONC

Concentration of the solution used for the command

.CYL

Cylinder volume of the exchange or dosing unit used
for the command

.DBL

Total duration for the processing of the command in s

.DRI

Current and/or last drift for drift correction in µL/min

.DSC

Time for processing all start conditions s

.DTI

Time for drift correction (time from the start of the
titration to the end of the command) in s

.EME

End measured value (measured value after processing
of the command) in the unit of the measured value