5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1106

■■■■■■■■

tiamo 2.3 (for Process Analysis)

(

)

B

i

N

i

c

c

z

U

E

U

+

+

=

log

)

0

(

The Nernst constant U

N

represents the theoretical slope s of the calibra-

tion function, divided by the charge z of the measuring ion. With anions it
has a negative sign, with cations a positive one.

The influence of the interfering ions is indicated as blank value concentra-
tion c

B

. A significant blank value reduces the lower measuring range and

therefore causes a deterioration of the detection limit.

For the calibration, there are thus three unknown quantities to be calcula-
ted, namely E(0), s and c

B

. In order to unambiguously calculate all parame-

ters, at least three standard measurements are thus necessary. A higher
number of standards increases the reliability of the mathematical calcula-
tion of the calibration function.

tiamo determines the calibration data for the concentration measurement
as for the pH calibration depending on the number of the standard solu-
tions used.

1 point calibration

s: the existing value is kept if available. Otherwise the theoretical

slope U

N

/ z is used.


c

B

: is set equal to 0.

2 point calibration


c

B

: is set equal to 0.