Metrohm tiamo 2.3 (process analysis) User Manual

Page 893

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.3 (for Process Analysis)

■■■■■■■■

877

Titrando, 855
Range

-20.000 ... 20.000

Default value

-20.000

Titrino
Range

-20.00 ... 20.00

Default value

-20.00

Lower hysteresis pH

Lower hysteresis of the measured value. If the measured value exceeds the
lower limit by this hysteresis value, then the event Measured value
lower limit ok
is triggered.

Titrando, 855
Range

0.000 ... 20.000

Default value

0.020

Upper limit pH

Upper limit of the measured value. If the measured value exceeds this
limit, then the event Measured value upper limit exceeded is trig-
gered.

Titrando, 855
Range

-20.000 ... 20.000

Default value

20.000

Titrino
Range

-20.00 ... 20.00

Default value

20.00

Upper hysteresis pH

The upper hysteresis of the measured value. If the measured value once
again falls below the upper limit by this hysteresis value, then the event
Measured value upper limit ok will be triggered.

Titrando, 855
Range

0.000 ... 20.000

Default value

0.020

Action

Selection of the event to be triggered if the upper or lower limit is excee-
ded: