3 properties - tower, Properties - tower # 0 – Metrohm tiamo 2.3 (process analysis) User Manual

Page 1576

Advertising
background image

7.13 864 Balance Sample Processor

■■■■■■■■■■■■■■■■■■■■■■

1560

■■■■■■■■

tiamo 2.3 (for Process Analysis)

Input

1000 characters

Default value

empty

7.13.3

Properties - Tower #

Tab: Configuration

Devices [Edit] Properties... Properties - 'Device type' -

'Device name'

Tower #

On the tab Tower the properties of the tower and of the robotic arm
connected to are displayed.

Tower parameters

Max. stroke path

Entry of the lowest permitted lift position. A lift height of 0 mm corre-
sponds to the "Home position", i.e. the lift is moved right to the top.

Range

0 ... 235 mm

Default value

235 mm

Min. beaker radius

Definition of the minimum radius which the beakers used on the rack
must have. If the lift is moved to the work position the beaker radii
defined in Rack table(see Chapter 6.12.2.1, page 1473) for general
Sample positions(see Chapter 6.12.3.2, page 1476) and Special beak-
ers
(see Chapter 6.12.3.4, page 1478) will be compared with the Min.
beaker radius
. If this minimum beaker radius is undercut then the run
will be stopped and an error message will appear. With off no check will
be made.

Selection

1.0 … 100.0 mm | off

Default value

off

Lift rate

Lift speed for manual operation.

Range

3 ... 25 mm/s

Default value

25 mm/s

Axial distance

Distance between the axis of rotation of the sample rack and swing axis of
the robotic arm.

Range

100.0 ... 300.0 mm

814 USB Sample Processor XL
Selection

166.0 mm

Default value

166.0 mm