5 Method

tiamo 2.3 (for Process Analysis)

■■■■■■■■

879

Lower hysteresis

Lower hysteresis of the dosing rate. If the dosing rate exceeds the lower
limit by this hysteresis value, then the event Dosing rate lower limit ok
is triggered.

Titrando, 855
Range

0.00 ... 166.00 mL/min

Default value

0.20 mL/min

Upper limit

Upper limit of the dosing rate. If the dosing rate exceeds this limit, the
event Dosing rate upper limit exceeded is triggered.

Titrando, 855
Range

0.00 ... 166.00 mL/min

Default value

166.00 mL/min

Titrino
Range

0.00 ... 150.00 mL/min

Default value

150.00 mL/min

Upper hysteresis

Upper hysteresis of the dosing rate. If the dosing rate once again falls
below the upper limit by this hysteresis value, then the event Dosing rate
upper limit ok
is triggered.

Titrando, 855
Range

0.00 ... 166.00 mL/min

Default value

0.20 mL/min

Action

Selection of the event to be triggered if the upper or lower limit is excee-
ded:

Selection

Cancel determination | Cancel command
(Titrando only)
| Wait for [Continue] | Wait for
limit ok
| none

Default value

none

Cancel determination
The running STAT command will be canceled, then the exit track (if
present) will be started and the determination will be finished.
Cancel command (Titrando only)
The running STAT command will be canceled, then the next com-
mand will be carried out.