2 starting determination series, Starting determination series 0 – Metrohm tiamo 2.3 (process analysis) User Manual

Page 1726

Advertising
background image

9.3 Determinations

■■■■■■■■■■■■■■■■■■■■■■

1710

■■■■■■■■

tiamo 2.3 (for Process Analysis)

3 If desired, enter Determination parameters.

4 In the Method list box select the method to be used for carrying out

the determination.

The subwindow Method shows the method loaded.

Note

In order to calculate statistical calculations for results, the Statis-
tics
option must be enabled in the START command of the
method and, in addition, the Statistics option must be enabled in
the CALC command for each desired result.

5 Activate the Statistics check box.

6 If desired, modify number of single determinations.

7 Enter sample data.

8 Click on [Start].

The single determination will be started. The subwindow Method
shows active tracks and commands, the subwindow Live display
live curves, measured values and messages.

Note

If the method contains a DATABASE command, then it is manda-
tory that this command be executed in all statistically linked deter-
minations. If this is not done, then the statistical results will be dis-
played incorrectly and the determinations cannot be reprocessed.

9.3.2

Starting determination series

How to proceed?

Determination series without statistics

1 Select program part Workplace.

2 Use the Run subwindow to select the Determination series tab.