5 Method

tiamo 2.3 (for Process Analysis)

■■■■■■■■

1221

If the option not defined is selected as Device name, any device types
or device groups which are able to execute the command can be selected,
independently of the devices in the device table.

Selection

Device types | 814 USB Sample Processor

Default value

814 USB Sample Processor

Stirrer

Stirrer

on | off (Default value: off)

Selection of the stirrer.

Titrando, 730, 846, 856, 867
Selection

1 | 2 | 3 | 4

Default value

1

778, 789
Selection

T1 | T2 (tower stirrer) | 1 … 3 (MSB stirrer)

Default value

T1

814, 815, 855, 864, 874
Selection

1 | 2 | 3

Default value

1

Titrino, 756, 831
Selection

1

Stirrer type

Display of the stirrer type. Unknown is displayed either with devices
unable to read out the stirrer type or if no stirrer is connected

Selection

801 Magnetic Stirrer | 802 Rod Stirrer | 803 Ti
Stand magnetic stirrer
| 804 Ti Stand rod stir-
rer
| unknown

Stirring rate

Setting the stirring rate or rotational speed. The algebraic sign of the stir-
ring rate changes the direction in which the stirring is done. The default
setting 8 corresponds to a rotational speed of 1000 rpm.

Titrando, 814, 815, 846, 855, 856, 867, 864, 874
Range

-15 ... 15

Default value

8