5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1240

■■■■■■■■

tiamo 2.3 (for Process Analysis)

Result monitoring

Result monitoring

on | off (Default value: off)

The question of whether the result is within the defined limit values is
monitored while calculating the result during the determination if this
option is enabled.

Lower limit

Lower limit for the result.

Input

10 digits

Upper limit

Upper limit for the result.

Input

10 digits

Message

Message

The message defined here can be put out to different targets if the upper
limit is exceeded or if the lower limit is not reached. The text editor for

entering or changing the message is opened with

or by double-click-

ing into the text field (see Chapter 2.5.2, page 90). The formula editor can
also be called up inside the text field (see Chapter 2.4, page 29).

Input

Text (unlimited)

Display message

on | off (Default value: on)

If this option is enabled, all active tracks are stopped and the message
defined in the text box will be displayed in the event that the limits are
exceeded. You can select whether you want to continue the stopped
tracks with [Continue] or - if defined - to execute the Action listed
below.

Record message

on | off (Default value: on)

If this option is enabled, the message defined in the text box will be docu-
mented in the determination if the limits are exceeded. The run will not be
stopped.