5 Method

tiamo 2.3 (for Process Analysis)

■■■■■■■■

769

User
Editing the individual titration parameters which affect the titration
rate.

Control

Note

Will be displayed only if the option User is selected in the selection list
Titration rate.

Control range

Dynamics defines the measured value range before the given endpoint. It
has a decisive influence on the titration rate and therefore on the accu-
racy. In the control range, dosing is controlled by the Min. rate. The
closer the endpoint, the slower the dosing until the Min. rate has been
reached. The larger the control range, the slower the titration. Outside the
range the dosing takes place with Max. rate (see Chapter 5.6.3.4.2,
page 693)
.

Titrando, 855
Range

0.01 ... 200.00 µA

Default value

40.0 (slow), 10.0 (optimal), 5.0 (fast) µA

Titrino
Range

0.1 ... 200.0 µA

Default value

40.0 (slow), 10.0 (optimal), 5.0 (fast) µA

Max. rate

Rate at which dosing is carried out outside of the control range. The
maximum dosing rate depends on the cylinder volume of the exchange
unit or dosing unit used.

Titrando, 855
Range

0.01 ... 166.00 mL/min

Default value

1.00 (slow), 10.00 (optimal), 166.00
(fast) mL/min

Titrino
Range

0.01 ... 150.00 mL/min

Default value

1.00 (slow), 10.00 (optimal), 150.00
(fast) mL/min