6.3 Configuration data

■■■■■■■■■■■■■■■■■■■■■■

1362

■■■■■■■■

tiamo 2.3 (for Process Analysis)

Min. slope

Minimum permitted slope of the pH electrode.

Range

-999.9 ... 999.9 %

Default value

95.0 %

Max. slope

Maximum permitted slope of the pH electrode.

Range

-999.9 ... 999.9 %

Default value

103.0 %

Max. response time

Maximum permitted response time of the pH electrode. The voltage
measured after three minutes in stirred solution is used as a comparison
value for determining the response time. The response time is the time
after which the measured voltage has reached this comparison value
except for ±1 mV.

Range

0 ... 999 s

Default value

60 s

Min. offset potential

Minimum permitted offset voltage (voltage at pH = 7.0). This value applies
for all evaluations.

Range

-999 ... 999 mV

Default value

-15 mV

Max. offset potential

Maximum permitted offset voltage (voltage at pH = 7.0). This value
applies for all evaluations.

Range

-999 ... 999 mV

Default value

15 mV

6.3.3.6.6

Data for specified electrode types

Tab: Configuration

Extras Templates Electrode types... [Proper-

ties]

Very good electrode / Good electrode / Usable electrode

The three predefined electrode types Standard, Gel and Non-aqueous
are always present in the table and be neither edited nor deleted. The fol-
lowing limit values are defined for the evaluation of the electrodes of
these electrode types: