3 digital in, 1 digital in - overview, Digital in – Metrohm tiamo 2.3 (process analysis) User Manual

Page 1307

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.3 (for Process Analysis)

■■■■■■■■

1291

Parameter 2

The meaning of Parameter 2 depends on the Type of the digital output
that has been set.

For Type = Stirrer:

0: Duration of the switch-on pulse fixed at 500 ms.

>0: For 100% stirrer output (Parameter 1 = 100), pulse duration in
10 ms
Example:
Parameter 1
= 100, Parameter 2 = 30: Pulse duration switched-on
300 ms, switched off 0 ms
Parameter 1 = 75, Parameter 2 = 30: Pulse duration switched-on
225 ms, switched off 75 ms

Range

-214783647 ... 214783647

Default value

0

5.6.11.3

DIGITAL IN

5.6.11.3.1

DIGITAL IN - Overview

Dialog window: Method

DIGITAL IN Properties... DIGITAL IN - 'Command

name'

Devices
This command can be executed with the IO Controller and with the Step-
ping Motor Controller.

Appearance
The command has the following appearance:

Mode of operation
The statuses of all digital inputs will be entered and saved in command
variables at the time of the command (syntax Command name.Variable
designation
)

Parameters
The parameters for the command DIGITAL IN are configured on the fol-
lowing 2 tabs:

General
Parameters for device.

Stop criteria
Parameters for defining the stop criteria.