Metrohm tiamo 2.3 (process analysis) User Manual

Page 539

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.3 (for Process Analysis)

■■■■■■■■

523

Titrino
Range

0.5 ... 999.0 mV/min

Default value

20.0 (slow), 50.0 (optimal), 80.0 (fast) mV/
min

Selection

off

off
Measured value acceptance will take place after the maximum waiting
time has elapsed.

Min. waiting time

The minimum waiting period is only important for drift-controlled meas-
urements. The measured value is not accepted until the minimum waiting
time has elapsed, even if the signal drift has already been reached. The
drift continues to be checked while the waiting time is elapsing.

Titrando, 855
Range

0 (slow, optimal, fast) ... 999999 s

Default value

0 (slow, optimal, fast) s

Max. waiting time

If signal drift has been switched off or has not yet been reached, then the
measured value will be accepted when the maximum waiting time has
elapsed. If the waiting time has not been newly entered then a waiting
time that is suitable for the drift will be calculated automatically according
to the following equation:

Titrando, 855
Range

0.1 ... 999999 s

Default value

38.0 (slow), 26.0 (optimal), 21.0 (fast) s

Titrino
Range

0.1 ... 9999 s

Default value

38.0 (slow), 26.0 (optimal), 21.0 (fast) s

Dosing of increments

Note

The following parameters will be displayed only if the option User is
selected in the selection list Titration rate.

Advertising