2 curve properties - x axis, X axis – Metrohm tiamo 2.3 (process analysis) User Manual

Page 398

Advertising
background image

4.8 Subwindow Curves

■■■■■■■■■■■■■■■■■■■■■■

382

■■■■■■■■

tiamo 2.3 (for Process Analysis)

Options
Options for the graphical display of curves.

Measuring point list
Parameters for the display of the measuring point list.

4.8.4.2

Curve properties - x axis

Tab: Database

Curves Properties curves # x axis

Parameters for the graphical display of the curves on the x axis.

Command type

Selection of the command type for which the curve properties are to be
defined. The curve properties defined for each command type are saved
per curve window and per client.

Selection

DET | MET | SET | MEAS | MEAS T/Flow | MEAS
TC Cond
| MEAS Ref | MEAS TMF | MEAS Spec
| MEAS Opt | CAL Spec | KFT | KFC | BRC | STAT
| DOS | TET

Default value

DET

Note

When the dialog window is opened, those Command type will be
selected per default which applies for the curve shown in the curve
window.

Autoscaling

on | off (Default value: on)

If this option is switched on then all axes in the curve window will be
scaled automatically. In this case the fields Start value and End value
cannot be edited.

Quantity

Selection of the quantity to be shown on the x axis.

For DET commands
Selection

Volume [mL] | Measured value | ERC | Time [s] |
Temperature [°C] | Calculated 1...3 | External
1...3

Default value

Volume [mL]

Advertising