5 kfc - titration parameters, 6 kfc - stop conditions, Titration parameters – Metrohm tiamo 2.3 (process analysis) User Manual

Page 851: Stop conditions

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.3 (for Process Analysis)

■■■■■■■■

835

5.6.3.6.5

KFC - Titration parameters

Tab: Method

KFC Properties... Titration parameters

Command name

Name of the command.

Input

25 characters

The parameters for the running of the titration are defined on this tab.

Extraction time

The titration is not stopped until the extraction time has elapsed (even
when the EP has already been reached). The entry of an extraction time
may be advisable, e.g. for the titration of sparingly soluble samples.

Range

0 ... 999999 s

Default value

0 s

Temperature

Manually entered titration temperature.

756, 831
Range

-170.0 ... 500.0 °C

Default value

25.0 °C

851, 852
Range

-20.0 ... 150.0 °C

Default value

25.0 °C

Time interval measuring point

Time interval for entering a measuring point in the measuring point list.

756, 831
Range

1 ... 999999 s

Default value

2 s

851, 852
Range

0.1 ... 999999.0 s

Default value

2.0 s

5.6.3.6.6

KFC - Stop conditions

Tab: Method

KFC Properties... Stop conditions

Command name

Name of the command.

Input

25 characters

Advertising