5 Method

tiamo 2.3 (for Process Analysis)

■■■■■■■■

853

Stabilizing time

Waiting period until the start of the titration during which the condition
requirement must be continuously fulfilled.

Range

0 ... 999 s

Default value

0 s

Automatic start after sample addition time

on | off (Default value: on)

If this option is activated, then the titration will be started automatically
after the sample addition time.

Sample addition time

Waiting time between stop of conditioning and start of titration during
which the sample can be added.

Range

0 ... 99 s

Default value

0 s

Manual start after sample addition with [Continue]

on | off (Default value: off)

If this option is activated, then the titration will be started manually after
the sample addition by pressing the [Continue] button in the message
window.

Only start titration by a start command from a SEND command

on | off (Default value: off)

If this option is activated, then the titration will not be started until the
Event message Start titration is sent to the command with a SEND
command.

Note

Conditioning must already be completed (= Condok) at the time the
SEND command is received, otherwise the command to start the titra-
tion will be ignored.

If this command is already started (e.g. via Automatic start after
sample addition
), the SEND command will not have any effect.