5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1150

■■■■■■■■

tiamo 2.3 (for Process Analysis)

Temperature

Temperature

The temperature at which calibration is carried out can be entered man-
ually. If a temperature sensor is connected and the Temperature meas-
urement
is set on the General/Hardware tab under Sensor to auto-
matic
or continuous, then the temperature will be measured continu-
ously. The value is used for temperature correction with the pH measure-
ment.

Range

-20.0 ... 150.0 °C

Default value

25.0 °C

5.6.6

Dosing commands

5.6.6.1

Dosing commands - Overview

Menu item: Method

Insert New command... Dosing

Commands for Dosing with dosing units and exchange units.

The following dosing commands can be selected:

ADD
Adding a predefined volume.

DOS pH
Controlled dosing with monitoring of measured value pH and tempera-
ture

DOS U
Controlled dosing with monitoring of measured value U and tempera-
ture

LQH
Extensive dosing possibilities with a Dosino.

PREP
Preparing a exchange or dosing unit.

EMPTY
Emptying a dosing unit.

RLS DOS
Releasing a dosing unit.

5.6.6.2

ADD

5.6.6.2.1

ADD - Overview

Dialog window: Method

ADD Properties... ADD - 'Command name'

Command for dosing a defined volume of a solution with an exchange
or dosing unit.

Devices
This command can be executed with the following devices: