3 cal meas opt - measuring parameters – Metrohm tiamo 2.3 (process analysis) User Manual

Page 1143

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.3 (for Process Analysis)

■■■■■■■■

1127

off
The temperature will not be measured. The temperature entered man-
ually under Temperature on the Measuring parameters tab will be
used.

Stirrer
The section Stirrer is only displayed for Device type = 089 Photome-
ter
.

Stirrer

Selection of the stirrer.

Selection

1 | off

Default value

off

Stirring rate

Setting the stirring rate.

Range

0 ... 15

Default value

3

Switch off automatically

on | off (Default value: on)

If this check box is activated, the stirrer will be switched off automatically
when the command has finished.

5.6.5.11.3

CAL MEAS Opt - Measuring parameters

Tab: Method

CAL MEAS Opt Properties... Measuring parameters

Command name

Name of the command.

Input

25 characters

The measuring parameters for the calibration are defined on this tab.

Parameters for the Avantes spectrometer

Wavelength

Wavelength at which the measured quantity (absorbance) is recorded.

Range

100.0 ... 2000.0 nm

Default value

400.0 nm

Blank value

Value by which the measured quantity (absorbance) is corrected. Here an
expression can also be entered through the formula editor (e.g. xx.EME