5 Method

tiamo 2.3 (for Process Analysis)

■■■■■■■■

905

Titrando, 855
Range

0.00 ... 166.00 mL/min

Default value

166.00 mL/min

Titrino
Range

0.00 ... 150.00 mL/min

Default value

150.00 mL/min

Upper hysteresis

Upper hysteresis of the dosing rate. If the dosing rate once again falls
below the upper limit by this hysteresis value, then the event Dosing rate
upper limit ok
is triggered.

Titrando, 855
Range

0.00 ... 166.00 mL/min

Default value

0.20 mL/min

Action

Selection of the event to be triggered if the upper or lower limit is excee-
ded:

Selection

Cancel determination | Cancel command
(Titrando, 855)
| Wait for [Continue] | Wait for
limit ok
| none

Default value

none

Cancel determination
The running STAT command will be canceled, then the exit track (if
present) will be started and the determination will be finished.
Cancel command (Titrando, 855)
The running STAT command will be canceled, then the next com-
mand will be carried out.
Wait for [Continue]
Reagent dosing in the running STAT command will be interrupted and
a message will be displayed. As soon as the monitored measured value
is again within the limits (including hysteresis) reagent dosing can be
resumed by pressing the [Continue] button in this message box.
Wait for limit ok
Reagent dosing in the current STAT command will be interrupted. As
soon as the monitored measured value is again within the limits
(including hysteresis), reagent dosing will be resumed automatically.
none
No action will be taken if limits are exceeded.