5 Method

tiamo 2.3 (for Process Analysis)

■■■■■■■■

703

5.6.3.4.3.4

SET pH - Control parameters

Tab: Method

SET pH Properties... Control parameters

Command name

Name of the command.

Input

25 characters

The control parameters can be set for each endpoint separately.

Endpoint 1

EP1 at pH

Measured value for the first endpoint.

Titrando, 855
Range

-20.000 ... 20.000

Selection

off

Default value

off

Titrino
Range

-20.00 ... 20.00

Selection

off

Default value

off

Titration rate

Titration rate

The three predefined sets of parameters can be selected for the titration
rate: slow, optimal and fast; the parameters for Control are not dis-
played. The User setting must be selected in order to be able to edit these
parameters.

Selection

slow | optimal | fast | User

Default value

optimal

slow
For titrations in which the finest details are also to be visible. This could
however also lead to an increase in noise, which could result in unwan-
ted EPs.
optimal
Parameter set for all standard titrations; optimized for the most fre-
quent applications.
fast
For less critical, fast titrations.
User
Editing the individual titration parameters which affect the titration
rate.