Metrohm tiamo 2.3 (process analysis) User Manual

Page 727

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.3 (for Process Analysis)

■■■■■■■■

711

Stabilizing time

Waiting period until the start of the titration during which the condition
requirement must be continuously fulfilled.

Range

0 ... 999 s

Default value

0 s

Automatic start after sample addition time

on | off (Default value: on)

If this option is activated, then the titration will be started automatically
after the sample addition time.

Sample addition time

Waiting time between stop of conditioning and start of titration during
which the sample can be added.

Range

0 ... 99 s

Default value

0 s

Manual start after sample addition with [Continue]

on | off (Default value: off)

If this option is activated, then the titration will be started manually after
the sample addition by pressing the [Continue] button in the message
window.

Only start titration by a start command from a SEND command

on | off (Default value: off)

If this option is activated, then the titration will not be started until the
Event message Start titration is sent to the command with a SEND
command.

Note

Conditioning must already be completed (= Condok) at the time the
SEND command is received, otherwise the command to start the
titration will be ignored.

Show measured value during conditioning

on | off (Default value: off)

If this option is enabled, then the measured value will be shown in the
subwindow Live display of the program part Workplace, in addition to
the drift and the time.