4 meas cond - evaluations – Metrohm tiamo 2.3 (process analysis) User Manual

Page 1036

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1020

■■■■■■■■

tiamo 2.3 (for Process Analysis)

Range

0.00 ... 9.99 %/°C

Default value

2.00 %/°C

Sample solution

The temperature compensation is carried out on the basis of temperature-
dependent coefficients and should be used for determinations where the
sample temperature is not constant. There are two possibilities:

It is possible to use a sample solution whose temperature coefficient
has been determined by a previous measurement (MEAS TC Cond).

For ground water, spring water and surface water, the temperature
coefficients according to standard ISO 7888:1985 (German version:
DIN EN 27888:1993) are stored in the system.

Selection

DIN | Table entries

Default value

DIN

5.6.4.10.4

MEAS Cond - Evaluations

Tab: Method

MEAS Cond Properties... Evaluations

Command name

Name of the command.

Input

25 characters

The following additional methods for evaluation of measurement curves
can be activated and defined on this tab.

Fixed endpoint evaluation

Fix endpoint evaluation

on | off (Default value: off)

If this option is activated, then the associated values for the other meas-
ured quantities will be interpolated with a fix measured quantity (Meas-
ured value
or Time) for the fix endpoint from the measuring point list.

Moves the selected line up (changes the sequence).

Moves the selected line down (changes the sequence).

[New]

Open the dialog window Fixed endpoint evaluation # to enter a new
fixed endpoint (see Chapter 5.6.4.10.6, page 1024).

Advertising