4 properties - limit values – Metrohm tiamo 2.3 (process analysis) User Manual

Page 1439

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

6 Configuration

tiamo 2.3 (for Process Analysis)

■■■■■■■■

1423

Cancel determination
The running determination will be automatically canceled. The follow-
ing message must be confirmed with [OK].

6.7.3.4

Properties - Limit values

Tab: Configuration

Sensors Edit Properties Limit values

Note

This tab is displayed only for sensors of the type pH electrode, ISE
electrode
or conductivity sensor.

Slope monitoring (for pH and ISE electrodes only)
on | off (Default value: off)

If this option is activated, then a check will be made to see whether the
Slope determined during calibration is within the defined limits. If it is
outside them a corresponding message appears and you can select
whether you will nevertheless accept the value or whether the previous
value is to be retained.

Lower limit

Lower limit value for the slope.

for pH electrode
Range

-2.147E+9 ... 2.147E+9 %

Default value

96.0 %

for ISE (+1)
Range

-2.147E+9 ... 2.147E+9 mV

Default value

55.0 mV

for ISE (-1)
Range

-2.147E+9 ... 2.147E+9 mV

Default value

-61.0 mV

for ISE (+2)
Range

-2.147E+9 ... 2.147E+9 mV

Default value

25.0 mV

for ISE (-2)
Range

-2.147E+9 ... 2.147E+9 mV

Default value

-31.0 mV

Advertising