Metrohm tiamo 2.3 (process analysis) User Manual

Page 519

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.3 (for Process Analysis)

■■■■■■■■

503

Fixed endpoint evaluation

Fix endpoint evaluation

on | off (Default value: off)

If this check box is activated, then the associated values for the other
measured quantities will be interpolated with a fixed measured quantity
(Measured value, Volume or Time) for the fix endpoint from the meas-
uring point list.

Moves the selected line up (changes the sequence).

Moves the selected line down (changes the sequence).

[New]

Open the dialog window Fixed endpoint evaluation # to enter a new
fixed endpoint (see Chapter 5.6.3.2.3.11, page 510).

[Properties]

Open the dialog window Fixed endpoint evaluation # to edit the fixed
endpoint selected in the table (see Chapter 5.6.3.2.3.11, page 510).

[Delete]

Delete the selected line.

pK/HNP evaluation

pK/HNP evaluation

on | off (Default value: off)

If this option is enabled, then the pK value is determined from the titration
curve which corresponds to the pH value at the half neutralization point
(see Chapter 5.6.3.10.1, page 924).

Minimum evaluation

Minimum evaluation

on | off (Default value: off)

If this option is enabled, then the associated volume, time and tempera-
ture for the minimum measured value will be interpolated from the meas-
uring point list.

Threshold value

The evaluation of the minimum begins as soon as the slope of the curve
exceeds the set threshold value.

Advertising