5 Method

tiamo 2.3 (for Process Analysis)

■■■■■■■■

1205

not defined
In this case, the device name must be assigned by the user at the start
of the method.

Device type

Display or selection of the device type. If a device is selected under Device
name
, then this Device type field can no longer be edited, and the
device type belonging to the selected device is displayed instead.

If the option not defined is selected as Device name, any device types
or device groups which are able to execute the command can be selected,
independently of the devices in the device table.

Selection

'Sample changer types' | 814 USB Sample Pro-
cessor

Default value

814 USB Sample Processor

Dosing device

Dosing device

Selection of the dosing device to be prepared. All the dosing device con-
nections which are possible with the selected device type are displayed.

778, 789, 814, 815, 855, 864, 874
Selection

1 | 2 | 3

Default value

1

730, 774
Selection

1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | 9 | 10 | 11 | 12

Default value

1

Solution

Input of a solution name or selection of a solution from the solutions
defined in the solution table. If intelligent dosing units are used, then a
check will be made in the method run to see whether the correct solution
has been attached to the selected dosing device. With non-intelligent dos-
ing units, only the cylinder volume is checked. At the start of the com-
mand, a check is made of the working life, the validity of the titer and the
GLP test interval for the selected solution.

Input

24 characters

Selection

'Solution name' | not defined

Default value

not defined

not defined
No tests will be carried out.