Metrohm tiamo 2.3 (process analysis) User Manual

Page 950

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

934

■■■■■■■■

tiamo 2.3 (for Process Analysis)

Evaluations
Defining further methods for evaluating measurement curves.

Additional measured values
Defining further measured values which can be saved as additional col-
umns in the measuring point list.

Command variables
The following command variables are generated in the method run by the
command MEAS pH and can be used in formulas under the designation
'Command name.Variable designation':

Designation

Description

.BP{x}.MEA

Measured value for the break point x (1…9) in the unit
of the measured value

.BP{x}.TEM

Temperature for the break point x (1…9) in °C

.BP{x}.TIM

Time for the break point x (1…9) in s

.BSY

Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY
; invalid (variable not available) = Command
has never been started

.DBL

Total duration for the processing of the command in s

.EME

End measured value (measured value after processing
of the command) in the unit of the measured value

.ENP

Electrode zero point of the sensor used for the com-
mand (dimensionless)

.ETE

End temperature (temperature following processing of
the command) in °C

.FIN

Command status; 1 = Command has ended at
least once
, 0, invalid (variable not available) = Com-
mand has never ended

.FP{x}.MEA

Measured value for the fixed endpoint x (1…9) in the
unit of the measured value

.FP{x}.TEM

Temperature for the fixed endpoint x (1…9) in °C

.FP{x}.TIM

Time in s until the fixed endpoint x (1…9) is reached

.IME

Initial measured value (measured value before start
conditions are processed) in the unit of the measured
value

.ITE

Initial temperature (temperature before start conditions
are processed) in °C

Advertising