5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

798

■■■■■■■■

tiamo 2.3 (for Process Analysis)

drift correction is the period between end of conditioning and end of
titration.

Selection

automatic | manual | off

Default value

off

automatic
The drift lastly determined during conditioning will be used as drift
value.
manual
The drift value defined under Drift value will be used.
off
No drift correction takes place.

Drift value

Drift value for drift correction. This parameter is only shown for Drift cor-
rection
= manual.

Range

0.0 ... 99.9 µL/min

Default value

0.0 µL/min

Stop volume

Determination stops when the specified volume has been dosed since the
start of the conditioning. The stop volume should be adjusted to the size
of the titration cell in order to prevent any overflow.

Titrando
Range

0.000000 ... 9999.99 mL

Default value

20.000 mL

Selection

off

Default value

off

Stop time

Determination stops when the specified time has elapsed since the start of
the conditioning.

Titrando
Range

1 ... 999999 s

Selection

off

Default value

off

Stabilizing time

Waiting period until the start of the titration during which the condition
requirement must be continuously fulfilled.

Range

0 ... 999 s

Default value

0 s