5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1076

■■■■■■■■

tiamo 2.3 (for Process Analysis)

Volume auxiliary solution

Volume of the auxiliary solution (e.g. buffer) to be added to the sample
before the first standard addition.

Range

0 ... 9999.999 mL

Default value

0 mL

Addition mode
Selection

Fixed volume increments | Variable volume
increments

Default value

Fixed volume increments

Fixed volume increments
If this option is enabled a fixed volume increment must be added for
each standard addition.
Variable volume increments
If this option is enabled the defined volume increment must be added
for each standard addition.

Addition volume

Size of the fixed addition volume (only when Addition mode = fixed
volume increments
).

Range

0.00000 ... 99999.9 mL

Default value

100 mL

Addition volume 1…19

Size of the individual addition volumes (only when Addition mode = var-
iable volume increments
).

Range

0.00000 ... 99999.9 mL

Default value

100 mL

Addition solution

Note

Displayed only for Solution = not defined.

Concentration

Concentration of addition solution.

Range

0.001 ... 999999.999

Default value

1.000