Metrohm tiamo 2.3 (process analysis) User Manual

Page 559

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.3 (for Process Analysis)

■■■■■■■■

543

not defined
No tests will be carried out.

Sensor

Measuring input

Selection of the measuring input the sensor is connected to.

Titrando (without 888)
Selection

1 | 2

Default value

1

855, 888
Selection

1

Default value

1

Sensor

Selection of a sensor of type Metal electrode from the sensors config-
ured in the sensor table. The calibration data for the sensor will be adop-
ted for the determination.

Selection

Sensor name | Metal electrode

Default value

Metal electrode

U(pol)

The polarization voltage is the potential applied to the polarizable elec-
trode during an amperometric measurement.

Titrando, 855
Range

–1250 ... 1250 mV (Increment: 25)

Default value

400 mV

Titrino
Range

–1270 ... 1270 mV (Increment: 10)

Default value

400 mV

Electrode check

on | off (Default value: off)

If this check box is enabled, then an electrode check will be carried out for
polarizable electrodes during the transition from an inactive normal status
to a measurement. A check is also made while doing so to ensure that the
electrode is properly connected and that no short-circuit is present.

Temperature measurement

Type of temperature measurement.

Advertising