INFICON IC/5 Thin Film Deposition Controller User Manual

Page 104

Advertising
background image

4 - 10

IP

N 07

4-

23

7A

E

IC/5 Operating Manual

Figure 4-7 Source Power Level Profile

MAXIMUM POWER . . . . . . . . . . . . . 0.0 to 99.9%

This parameter is used to set the maximum permissible % power level. The
control voltage output will not exceed this limit. Values range from 0.0 to 99.9%.
The default value is 90%.

Power Ramps (Pre deposit states)

The next six parameters—Soak Power 1, Rise Time 1, Soak Time 1, and Soak
Power 2, Rise Time 2 and Soak Time 2—are used to define two power ramps
provided to precondition materials.

NOTE: The first power ramp consisting of Rise 1 and Soak 1 is skipped if the

previous layer had the same source, same crucible and its idle power is
nonzero.

SOAK POWER 1 . . . . . . . . . . . . . . . 0.0 to 99.9%

This parameter is usually set to the power level at which the source material
just begins to melt. The instrument ramps the power level from zero to Soak
Power 1 linearly over the time period Rise Time 1. Values range from 0.0 to
99.9%. The default value is 0.

RISE TIME 1 . . . . . . . . . . . . . . . . . . 00:00 to 99:59 min:sec

This parameter provides the time period over which the source power is
ramped from 0 to Soak Power 1. Values range from 00:00 to 99:59
minutes:seconds. The default values is 00:00.

SOAK TIME 1 . . . . . . . . . . . . . . . . . 00:00 to 99:59 min:sec

This parameter provides the time period for which the instrument holds at Soak
Power 1. Values range from 00:00 to 99:59 minutes. The default value is 00:00.