IC/5 Operating Manual

Figure 12-3 AutoTune Parameters Display

12.6.1 AutoTune Parameters

Refer to Sections

section 2.1.7 on page 2-11

and

section 12.6.3 on page 12-14

for

a description of AutoTune.

MATERIAL INDEX. . . . . . . . . . . . . . 1 to 24

This parameter is used to select the Material (from the MATERIAL
DIRECTORY) that is to be "AutoTuned". Values range from 1 to 24. The default
value is 1.

DESIRED RATE . . . . . . . . . . . . . . . 0.1 to 999 Å/s

For a Complete Tune, this parameter determines the Rate for which the
AutoTuned control loop parameters will be calculated. For Quick Tune this
parameter is ignored. Values range from 0.1 to 999 Å/s. The default value is
10.0 Å/s.

MAXIMUM RATE. . . . . . . . . . . . . . . 0.2 to 999 Å/s

This parameter is used to set the maximum deposition Rate allowed during an
AutoTune session. The events which occur upon exceeding this value are
dependent upon whether a Quick Tune or a Complete Tune is designated. The
range of values permitted for this parameter is dependent on the value chosen
for the Desired Rate. The maximum allowed value will always be 999 Å/s. The
minimum allowed value is 2 times the Desired Rate value.

QUICK TUNE. . . . . . . . . . . . . . . . . . YES/NO

This parameter is used to choose between a Quick Tune and a Complete Tune.
Permissible values are YES and NO. A value of YES selects Quick Tune. The
default value is YES.

MATERIAL NAME

This is a read-only parameter that identifies the material being deposited.