7.7.4 UPDATE Commands

Update commands are used to change a specified parameter to a new value or
condition. Some parameters are order dependent in that they cannot be updated
prior to certain other parameters. For example, in the IC/5 two layers must be
defined before specifying the codeposition of those layers.

There is a specific update command for each parameter group. Each command
has one to four arguments starting with and separated by a space. The Update
command variations include:

7.7.4.1 Update Material Parameters

For each command format the symbol

þ

represents a blank space.

The format is:

UM

þ

param#

þ

material#

þ

value

This command changes the value of a material parameter for a specified material.
Refer to

Chapter 4, Material Set Up

, for parameter value ranges.

Table 7-11 Update Commands

command

parameter

UM

Update Material Parameters

UP

Update Process Parameters

UU

Update Utility Parameters

US

Update Sensor Parameters

UC

Update Source Control Parameters

UT

Update Output Relay Type

UL

Update Logic Statement

UO

Update alphanumeric name of hardware output

UI

Update alphanumeric name of hardware input

UN

Update user messages

Table 7-12 Update Material Parameters

param# Material

Parameter

1

Density

2

Z-Ratio

3

Source

4

Control Loop

5

Process Gain

6

Primary Time Constant

7

System Dead Time

8

Recorder Output

9

Recorder Function

10

Master Tooling