INFICON IC/5 Thin Film Deposition Controller User Manual

Page 87

Advertising
background image

3 - 35

IP

N 07

4-

23

7A

E

IC/5 Operating Manual

3

Select the Crucible Output Type as normally open (NO), or normally closed
(NC).

Example:

Number of crucibles

= 4

Crucible output

= 6

Crucible output type

= NO

For this example, wiring to the controller is based on

Table 3-4

. Only relays 6

and 7 are needed to encode the four possible positions.

NOTE: If the crucible output type were normally closed (NC),

Table 3-4

would

need to be modified by exchanging open and closed.

4

Determine whether Turret Feedback is desired. This allows the turret position
controller to stop further instrument processing until the requested turret
position is satisfied. If chosen, a turret input must be connected to the turret
position controller’s feedback signal.

If Turret Feedback is not chosen, program a Turret Delay Time which allows an
adequate time for positioning to take place. Once the delay time has expired,
instrument state processing continues.

5

The selection of a particular crucible for a layer is defined in the process
directory.

6

Select the process to modify.

7

Program the particular “Crucible Number” for each layer.

Example:

Layer 1:

Crucible Number

= 1

Layer 2:

Crucible Number

= 2

Layer 4:

Crucible Number

= 4

Table 3-4 Wiring To The Controller

Crucible
Position

Contact Status

Relay #6

Relay #7

1

Open

Open

2

Closed

Open

3

Open

Closed

4

Closed

Closed