2 update process parameters – INFICON IC/5 Thin Film Deposition Controller User Manual

Page 163

Advertising
background image

7 - 21

IP

N 07

4-

23

7A

E

IC/5 Operating Manual

7.7.4.2 Update Process Parameters

For each command format the symbol

þ

represents a blank space.

The format is:

UP

þ

param#

þ

process#

þ

layer#

þ

value

This command changes the value of a process parameter for a specific process
and layer. Refer to

Chapter 5, Process Set-Up

, for parameter value ranges.

NOTE: Rate, New Rate 1, New Rate 2 values can be input to the hundredth

position via communications. Deposition will be controlled to the hundredth
place, even though the front panel display and input only read to the tenth’s
place.

Table 7-13 Update Process Parameters

param# Process

Parameter

0

Material Index

1

Aggregate Rate

2

Final Thickness

3

Thickness Limit

4

Time Limit

5

Co-Deposition Value is upper or lower case YES or NO

6

Ratio Control

7

Rate Watch Time

8

Rate Watch Accuracy

9

Crucible

10

New Rate 1

11

Start Ramp 1

12

Ramp Time 1

13

New Rate 2

14

Start Ramp 2

15

Ramp Time 2