IC/5 Operating Manual

CRUCIBLE OUTPUT. . . . . . . . . . . . 0 to 37

This parameter designates which of the output relays are to be used as
Crucible Output relays. Values range from 0 to 37. The default is 0 and
indicates the Crucible Output is inactive. The value entered into this parameter
indicates which of the output relays begins the sequence of outputs used as
crucible control outputs. For example, a 4 entered into the Number of Crucibles
parameter and a 1 entered into the Crucible Output parameter will designate
output relays 1 and 2 as crucible control outputs with the least significant bit of
the binary coding in relay output 1. An 8 entered into the Number of Crucibles
parameter and a 1 entered into the Crucible Output parameter will designate
output relays 1, 2, and 3 as crucible control outputs with the least significant bit
of the binary coding in relay output 1.

CRUCIBLE OUTPUT TYPE . . . . . . 0/1

This parameter designates the normal state of the relay contacts for the
Crucible Output relays. Permissible values are 0 and 1. 0 indicates an Normally
Open relay, 1 indicates a Normally Closed relay.

TURRET FEEDBACK . . . . . . . . . . . YES/NO

Some turret source indexers provide feedback to signify when the turret is in
the proper position. This parameter allows the IC/5 to accept this input and
respond accordingly. Parameter entry may be either YES or NO. A YES
indicates that Turret Feedback is expected and the Turret Input parameter is
displayed on the screen. A NO indicates there is no Turret Feedback
forthcoming and the Turret Delay parameter is displayed. The default value is
NO. See the Turret Input and Turret Delay descriptions below.

NOTE: If the turret feedback is set to YES and the Process is STARTed, the

IC/5 will proceed to the source switch state. If the turret input is not
received within 60 seconds the IC/5 will enter the STOP state.

TURRET INPUT. . . . . . . . . . . . . . . . 0 to 28

This parameter designates which of the 28 inputs is to be the Turret Feedback
input. Values range from 0 to 28. Values 1 through 28 correspond to the
appropriate input. Value 0 indicates that an input for turret feedback is not used.
The default value is 0. However, an input must be assigned, or the process will
never proceed past the source switching phase of the process. If a value other
than 0 is entered, then the input chosen as the Turret Input is designated as
such on the I/O Map display. Input lines are activated with a contact closure to
ground.