TURRET DELAY . . . . . . . . . . . . . . . 1 to 60 sec
If Turret Feedback is not used a timer will be set to allow the source crucible
time to rotate into position. This parameter sets the amount of time the Process
will remain in the Source Switch state waiting for the crucible to rotate before
proceeding to the Rise 1 state. Permissible values are 1 to 60 seconds. The
default value is 5 seconds.
CAUTION
The IC/5 will proceed to the Rise 1 state after the Turret
Delay time expires. This could result in damage to your
equipment if the crucible is not in position after
expiration of the Turret Delay time.
Figure 8-4 Sensor Parameter Editing