5 source maintenance display – INFICON IC/5 Thin Film Deposition Controller User Manual

Page 70

Advertising
background image

3 - 18

IP

N 07

4-

23

7A

E

IC/5 Operating Manual

3.3.5 Source Maintenance Display

Figure 3-10 Source Maintenance Display

3.3.5.1 Function Key Selection Choices for the

Source Maintenance Display

F1 SWITCH CRUCIBLE

To switch the source turret to the crucible position designated by the crucible
parameter, press function key F1.

F2 TOGGLE SENSOR SHUTTERS

To activate the shutter relay for the sensors associated with the designated
material, press function key F2. Pressing this key a second time deactivates the
sensor shutter relays.

F3 TOGGLE SOURCE

To activate the shutter relay for the source associated with the designated
material, press function key F3. A second pressing of this key deactivates the
source shutter relay.

F4 START MANUAL POWER

To enable the hand-held controller to effect power changes to the source output
associated with the chosen material, press function key F4. Pressing this key
a second time exits manual power.
NOTE: The MANUAL power level is automatically set to zero % power when

exiting Manual Power while in source maintenance. This is different
from executing a process. When executing a process, pressing the
Manual Power key a second time will place the instrument into the
Deposit State.

F5 AUTOTUNE

To activate the Autotune displays, press function key F5. See

section 12.6 on

page 12-11

for a description of Autotune and Autotune displays.

F6 MAIN/DIAG

To return to the Maintenance/Diagnostics display, press function key F6.

Advertising