2 autotune manual display – INFICON IC/5 Thin Film Deposition Controller User Manual

Page 259

Advertising
background image

12 - 13

IP

N 07

4-

23

7A

E

IC/5 Operating Manual

CONTROL LOOP

This is a read-only parameter that indicates the type of control loop specified
during material definition. A 0 indicates the Non-PID control loop is used. A 1
indicates a PI control loop is used, and a 2 indicates a PID control loop is being
used. See

section 4.3 on page 4-5

and

section 2.1.7 on page 2-11

for more

information about the control loop.

PROCESS GAIN

This is a read-only parameter that represents the rate of change of % source
power for a given rate deviation.

PRIMARY TIME CONSTANT

This is a read-only parameter that represents the source's time constant. It is
valid only for a PI or PID control loop.

SYSTEM DEAD TIME

This is a read-only parameter that represents the source's lag time. It is valid
only for a PI or PID control loop.

12.6.2 AutoTune Manual Display

The AutoTune Manual display is entered from the AutoTune Parameters display by
selecting START AUTOTUNE (F6).

The AutoTune Manual display is similar to the general Operate display with the
exception that Thickness and Process information will be absent. The AutoTune
Manual display automatically places the instrument into the MANUAL state. Just
like the normal MANUAL state, the power level can be adjusted via the Hand-Held
Controller and the source shutter relay is active.