INFICON IC/5 Thin Film Deposition Controller User Manual

Page 83

Advertising
background image

3 - 31

IP

N 07

4-

23

7A

E

IC/5 Operating Manual

12. RATE RAMP TIME 2

Rate control, desired rate
changing. [New Rate 2, Start Ramp
2, Ramp Time 2]

Active

Active

13. RATEWATCHER

(SAMPLE)

Rate control. [RateWatch
Accuracy]

Active

Active

14. RATEWATCHER

(HOLD)

Constant power, based on last
sample’s average power.
[RateWatch Time]

Active

Inactive

15. MANUAL

Source power controlled by
hand-held controller.

Active

Active

16. TIME-POWER

Crystal failed; source maintained at
average control power prior to
crystal failure. [Time Pwr Y]

Active

Active

NOTE: 17 through 20 are Post-Deposit states.

17. FEED RAMP

Source changing to Feed Power.
[Feed Power, Feed Ramp Time]

Inactive

Inactive

18. FEED

Source maintained at Feed Power
level. [Feed Time]

Inactive

Inactive

19. IDLE RAMP

Source changing to Idle Power.
[Idle Ramp Time, Idle Power]

Inactive

Inactive

20a.IDLE POWER (=0%)

Source maintained at zero power;
will accept a START command.

Inactive

Inactive

20b.IDLE POWER (>0%)

Source resting at Idle Power; will
accept a START command.

Inactive

Inactive

21. STOP

Source output set to zero power.
The display is frozen at the last rate
and thickness values; will not
accept a START or RESET
command.

Inactive

Inactive

NOTE: In the STOP state the instrument will accept a START
provided a Crystal Failure has not occurred for the sensors
used in the layer being started.

Table 3-3 State Descriptions (continued)

STATE

CONDITION

RELAY CONTACT

STATUS

NOTE: 1 through 7 are Pre-Deposit states.

Source
Shutter

Sensor
Shutter

Advertising