9 error messages in i/o logic statement set-up – INFICON IC/5 Thin Film Deposition Controller User Manual

Page 141

Advertising
background image

6 - 21

IP

N 07

4-

23

7A

E

IC/5 Operating Manual

6.9 Error Messages in I/O

Logic Statement Set-Up

CANNOT TOGGLE INPUT

Output types can be toggled between Normally Open and Normally Closed. An
Input does not have a type function and cannot be toggled.

EDIT IN PROGRESS

The I/O Map or the User Messages display cannot be exited while an edit name
or edit message operation is occurring.

ILLEGAL INPUT - VALUE TO LARGE
ILLEGAL INPUT - VALUE TOO SMALL

Parameter entry is out of range. Allowable values will vary according to your
unit configuration or the parameter being defined. Press Clear to delete the
value and enter again.

INPUT IN USE

The input line in question has already been reserved for use.

INVALID EQUATION

Strings must have connectors and required numerics before exiting the Logic
Statement

MAXIMUM TERMS

No more than five terms are allowed in any one Event or Action string.

MUST EDIT NAME FIRST

A save name or save message operation was requested without first having
edited the name or message.

OUTPUT IN USE

Indicates the output in question is already reserved for use in Sensor or Source
set-up or in one of the Logic Statements.

Advertising