Clearing alarm trace-back data – Yaskawa FSP Amplifier User Manual

Page 216

Advertising
background image

FSP Amplifier User’s Manual

Chapter 7: Using the Panel Operator

7-25

7.2.5. Clearing Alarm Trace-back Data

This procedure clears the alarm history, which stores alarms generated in
the servo amplifier. After clearing, each alarm in the alarm history is set to
A.- -, which is not an alarm code. Refer to 7.2.1 Operation in Alarm Trace-
back Mode
for details.
Follow the procedure below to clear the alarm trace-back data.

1. Press the MODE/SET key to select the Auxiliary Function mode.

2. Press the Up Arrow or Down Arrow key to select the function Fn006.

3. Press the DATA/SHIFT key for a minimum of one second, and the

display will be as shown below.

4. Press the MODE/SET key to clear the alarm trace-back data. The

following display will flash for one second, and after the alarm trace-
back data is cleared, the display will return to show as above in step 3.

Flashing for

one second

5. Press the DATA/SHIFT key for a minimum of one second to return to

the parameter code display.

This completes the alarm trace-back data clearing procedure.

Advertising