The HAPSITE will re-run the cleanout as part of its preparation.

5

If the failure box appears again, check the concentrator to verify that it is not
cracked or chipped. Also, try re-installing the concentrator to ensure that it is
properly seated.

3.1.4 Quick Reference SOP — Heat-up and Tune

CAUTION

Do not open the front panel in a wet or contaminated area.

1

Insert the internal standard and carrier gas canisters.

2

Insert a charged battery.

3

Connect the AC to DC power converter power supply.

4

Verify that the appropriate sample configuration (i.e., concentrator) is installed.

5

Press the POWER button on the front panel.

6

The HAPSITE will heat up the necessary components and perform AutoTune.
A prompt to run SURVEY or ANALYZE will appear when the HAPSITE is ready
to run a sample. The process for heating and tuning takes approximately 20
minutes.

7

If the default method is not the desired method, touch STOP PREPARE.

8

Touch SELECT METHOD. Highlight the desired method. Touch Select.

NOTE: If the wireless connection to the Laptop is to be used, see

Chapter 4,

Communications and Touch Screen Options

.

NOTE: When the SYSTEM READY message is displayed, touch either RUN

SURVEY or RUN ANALYZE. If using the push buttons, push SURVEY
RUN
or ANALYZE RUN.