16 - 10

IP

N 07

4-

47

2-

P1

C

HAPSITE Smart Plus Operating Manual

4

With each end of the Probe in a separate hand, alternately move each ends of
the Probe Line up and down until methanol comes out of either end of the
Probe. Repeat this a couple of times so the methanol can dissolve any
contaminants in the Probe Line. See

Figure 16-2

.

Figure 16-2 Moving Methanol Around in Probe and Probe Line

5

Empty the methanol remaining in the Probe Line by tipping one end of the
Probe Line downward until all of the methanol drains from it. See

Figure 16-3

.

Figure 16-3 Empty Methanol from Probe and Probe Line

6

Blow out the Probe Line with Nitrogen to remove any residual methanol that
may be left in the Probe Line. See

Figure 16-4

.

WARNING

Point both ends of Probe away from you when blowing
out the Probe Line with Nitrogen. Continue to wear
appropriate Personal Protection Equipment (PPE) while
blowing out the Probe Line.