HAPSITE Smart Plus Operating Manual

Figure 14-6 Figure 6-3. Multi-Point Internal Standard Calibration

14.13 Maintenance

Periodic cleaning is the only maintenance required for the SituProbe accessory.
The Viton® o-ring, between the sample collection tube and the purge head, should
be checked and cleaned or replaced as necessary.

14.13.1 Surface Cleaning

Probe head surfaces will eventually become fouled with aquatic and bacterial
growth from long-term contact with water samples. Surface coating of the
conductivity sensors can result in sensing errors and method termination. Coatings
on the sample collection tube and probe head may trap VOCs and contribute to
carryover. The sample collection tube can be unscrewed from the purge head and
cleaned with soap and water, followed by thorough rinsing with clean water.

NOTE: To remove the sample collection tube, hold the probe head close to the

sample collection tube. Unscrew only the sample collection tube. Do not
unscrew the probe head. See

Figure 14-7

.

The o-ring can be cleaned with a lint-free cloth and clean water. Gently wipe the
electrodes and probe head surfaces with a wet lint-free cloth. Rinse thoroughly with
clean water. Regular cleaning is recommended as a preventive measure.