HAPSITE Smart Plus Operating Manual
2
Remove the new NEG Pump and aluminum gasket from the shipping
container. See
Figure 16-27
.
Figure 16-27 New NEG Pump and Aluminum Gasket
3
Clean the aluminum gasket with methanol on a lint free wipe. Be careful to
handle the aluminum gasket by the outer band and keep the knife-edges on
the inner ring from getting nicked. See
Figure 16-28
.
Figure 16-28 Cleaning Aluminum Gasket