3 cleaning the hss wells, Warning – INFICON HAPSITE Smart Plus Chemical Identification System User Manual

Page 453

Advertising
background image

13 - 15

IP

N 07

4-

47

2-

P1

C

HAPSITE Smart Plus Operating Manual

Figure 13-9 Flush Switch Inside The Headspace Instrument

13.3.3 Cleaning the HSS Wells

The heating block is constructed of aluminum and is normally easy to keep clean.
Routine cleaning of the wells is not usually necessary, but debris and liquids may
be removed easily.

WARNING

Be careful to avoid burns if the surfaces are hot. Surfaces
may still be hot, even if the HSS has been turned off.

WARNING

Be careful to avoid injury from the sharp sampling
needle.

The four wells may be cleaned, as necessary, using a damp paper towel or cloth.
Cleaning should be performed after the heating block has been allowed to cool to
ambient temperature or a reasonable temperature to avoid injury.

HINT: When cleaning the wells, position the needle assembly inside a well not

currently being cleaned. Use this as a precaution against possible injury
from the sharp needle point of the assembly.

To remove liquids, roll a paper towel or cloth into a tube shape which will fit easily
into the well, insert, twist and remove.