HAPSITE Smart Plus Operating Manual

10

Select Custom. Type 60 in the Start Field and 90 in the End Field. This will
enable the analyst to scan for both the 69 and 83 masses. (These are the
predominant masses in a can of MicroDuster that contains
1,1,1,2-tetrafluorefloroethane, the compound being used to check for leaks.)
See

Figure 16-43

.

Figure 16-43 Checking For Leaks